Abstract

Subclinical leaflet thrombosis has been increasingly observed in patients undergoing transcatheter aortic valve replacement. Intra-annular transcatheter aortic valves (TAVs) have a larger neo-sinus volume than supra-annular devices and are potentially at a higher risk of hypoattenuated leaflet thickening (HALT). However, clinical data from randomized clinical trials have shown that approximately one-third of patients undergoing TAVR with intra- or supra-annular devices develop HALT in 1year. The findings point to the potential role of leaflet design in developing HALT. The study aimed to systematically investigate leaflet kinematics of a supra-annular TAV, Medtronic CoreValve, and determine regions of blood stasis. Fluid-solid interaction simulations demonstrated the limited movement of CoreValve leaflets in the lower belly region that created regions of blood stasis on the surface of the leaflets. The findings provide insights into potential improvements in leaflet design in the next generation of TAVs to reduce the risk of HALT and leaflet immobility.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.