Abstract
This study compared the tissue heating produced by 45kHz and 1MHz ultrasound machines used subaqueously in metal and in plastic basins. An intervention study with cross-over design was used. The setting was a laboratory. Two cross-sections of nonliving pig tissue were used with thermocouples positioned, in the tissues, at different distances along the ultrasound beam axis. Specimens were each exposed subaqueously to ultrasound produced by a 45kHz machine and a 1MHz machine. Both metal and plastic treatment basins were used (2 machines × 2 basin types). Thermocouples measured temperature variations for 30 minutes with each machine on and 10 minutes with it off. Temperature increases at different tissue depths varied with both the ultrasound machine and the type of basin used. The 45kHz frequency ultrasound machine produced a maximum temperature increase of 0.4°C and an initial rate of heating of 0.05°C per minute. The 1MHz machine consistently produced greater temperature increases. These were further increased by 2.65°C (average) if a metal rather than plastic basin was used. The 45kHz machine, operated at maximum output, produced little heating of the irradiated tissue. By contrast, the 1MHz ultrasound unit produced appreciable heating to almost the full depth of the tissue irradiated. Use of a metal basin, rather than plastic, resulted in a markedly higher maximum temperature increase and a significantly greater initial rate of heating.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.