Abstract
We report here a low-cost electrochemical immunoarray with unprecedented sensitivity in the sub-zeptomole range with up to 5 log-decades dynamic range for accurate, multiplexed protein determinations. The microfluidic array features eight carbon sensors coated with a dense layer of 5 nm gold-nanoparticles derivatized with primary antibodies. Analyte proteins are captured by secondary antibody-poly-HPR (horseradish peroxidase) bioconjugates containing 400 HRP enzyme labels, with amplified amperometric peaks developed using H2O2 activator and hydroquinone mediator. Prostate cancer biomarkers prostate specific antigen (PSA), vascular endothelial growth factor-D (VEGF-D), ETS-related gene protein (ERG), and insulin-like growth factor-1 (IGF-1) were measured simultaneously with sub-fg/mL LODs (0.08-0.22 zmol). These proteins were determined in serum of postprostatectomy cancer patients which had much lower levels than prostate cancer patients without surgery. This immunoassay protocol makes thousands of low-abundance proteins accessible to quantitative measurements down to zeptomole levels.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.