Abstract

Abstract Microcharacterization of defects is greatly facilitated with improvement in resolution. By improving temporal coherence (chromatic aberration) and compensating spatial incoherence, we have achieved the goal of the 1 Å microscope (OÅM) project at the US Department of Energy's National Center for Electron Microscopy by extending the limits of high-resolution transmission electron microscopy to sub-ångström levels. The OÅM combines focal-series image-processing software with a modified 300 keV electron microscope equipped with a highly coherent field emission electron gun. By operating at an ‘α-null’ value of underfocus in order to minimize the effects of spatial incoherence, and by reducing the OÅM's electron-gun extraction voltage to improve temporal coherence, we are able to transfer information below 0.8 Å. In a test specimen of silicon viewed in the [112] orientation in the OÅM we are able to ‘see’ atoms separated by only 0.78 Å. Sub-ångström resolution at this level offers the materials researcher an effective tool for the characterization of defects with unprecedented precision.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.