Abstract
Purpose: To validate the accuracy of missing tissue compensators in conventional techniques calculated by the Pinnacle 3D treatment planning system version 7.6c. Method and Materials: For the Pinnacle version 7.6c, the density in g/cm3 of lead material and the distance from the source to the top of the compensator tray were specified. Lead sheet thicknesses were entered manually pixel by pixel to create the compensators dimension and position corresponding to the missing surface until the isodose profiles were uniform. In solid water phantoms, the lead sheet missing tissue compensators were designed in the conventional techniques at depth 5 and 10 cm. The thicknesses of 1, 3, 5 and 7 cm of step solid water phantoms was used to simulate as a missing tissue. Kodak X‐Omat V film was used to verify the ID dose profiles. OmniPro‐I'mRT software analysis was employed for the dose comparison between measurement and calculation. Results: The study showed that, the missing tissue thicknesses of 1, 3, 5 and 7 cm need to be compensated using lead sheets thicknesses of 0.65, 2.06, 3.62 and 5.42 mm, respectively. The difference of all dose profiles between the Pinnacle 3D treatment planning and the film measurement are within 3% which agree in the criteria of TRS 430 protocol. Conclusions: The missing tissue compensators using the Pinnacle 3D treatment planning version 7.6c establish the uniform dose distributions for 1D profile comparison which can be applied in clinical situations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.