Abstract

Purpose: The goal of our research is to determine the viability of optically stimulated luminescent (OSL) dosimeters in verifying the reported CTDIvol from CTscanners.Methods & Materials: We used a Siemens Somatom Sensation 16 slice scanner, a CTDI phantom, aluminum oxide (Al2O3:C) OSLnanoDot dosimeters read by the Landauer InLight® microStar system, and an ion chamber with an RTI Barracuda control unit. Exposures were made with routine adult head and abdomen protocols, since these have specific ACR recommendations, for the full length of the phantom. Dosimeters were placed in the middle of the phantom at the 12, 3, 6, and 9 clock positions, as well as the center. Results: We found the doses measured by the nanoDots to be higher than what was reported by the scanner, but closer than what was measured by the ion chamber, which was lower. Conclusion: We plan further study of the ability to use nanoDots with other protocols where the CTDI phantom could be used.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.