Abstract

Purpose:To compare the dosimetric difference in forward intensity modulation radiotherapy(fIMRT) and field in field IMRT (FIF IMRT)planning for breast cancer.Methods:Ten patients received radiotherapy are selected.For each patient,two treatment plans(fIMRT and FIF IMRT) were designed with Varian Eclipse ver11.0 treatment planning system.Evaluate the dose parameters of targets, organs at risk (OAR), monitor units and treatment time, using dose‐volume histogram (DVH).Results:There were no significant difference were found in conformal index (CI), homogeneity index (HI) of PTV,V5,V10,V20,V30,V40,V50 of heart, lung and monitor unit(MU)(P>0.05).The differences were significant in the treatment time(fIMRT=8.3min,FIF IMRT=2.5,p<0.05).Conclusion:FIF IMRT is equal to fIMRT in dosimetril evaluation. Due to much less delivery time,FIF IMRT is an efficient technique in treating patients by reduceing the uncomfortable influnce which could effect the treatment.