Abstract

Purpose: Evaluating timing differences between LINAC beam ON/OFF and the estimation of tumor positioning using gating systems is essential for establishing confidence when treating with gating during radiotherapy, and is an annual requirement of TG-142. Temporal discrepancies between the trajectories of external marker surrogates and beam delivery may vary depending upon the type of external marker motion, which is quantified in this work for several trajectories. Methods: A precise robotic 3D motion stage performed several trajectories typically used for gating phantoms, including sinusoidal and Lujan-type motion; a commercial respiratory motion simulator was also employed. The true motions were monitored using variable resistors. The beam ON/OFF was controlled separately by two RPM (Varian) systems, an integrated version delivered by a Varian Truebeam LINAC and version 1.6 delivered by a Varian Trilogy, and measured using a diode. The resistor and diode signals were read by a multichannel digital oscilloscope, and timing differences between beam ON/OFF as measured by the diode and the phantom motion were determined using a peak detection algorithm. Results: Timing differences between beam ON/OFF and 3D stage motion peaks (diode—true motion timing) were computed to be 79.4 & 57.7ms for sinusoidal motion and 109.1 & 63.6ms for Lujan-type motion on the Truebeam LINAC, for beam ON and OFF, respectively. Timing differences for the Trilogy LINAC were 34.4 & 55.2ms for the sinusoidal motion and 29.0 & 26.3ms for the Lujan-type motion, for beam ON and OFF, respectively. With the commercial motion simulator, the timing differences were found to be −9ms and −78ms for beam ON/OFF, respectively, with the Truebeam, and −97.6ms and −60.9ms for beam ON/OFF, respectively, with the Trilogy. Conclusion: Setup-dependent temporal lags were found using this methodology. These discrepancies have the potential to influence quality assurance on gating systems and ultimately the treatment dose during radiotherapy. NIH National Institute of Biomedical Imaging and Bioengineering grant T32 EB002103-21

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.