Abstract

Purpose:The purpose of this work is to develop and verify our quality assurance (QA) procedures to ensure the safety and efficacy of MR‐guided focused ultrasound (MRgFUS) treatment of bone metastases.Methods:A practical QA program was developed. Monthly and daily QA (DQA) procedures were performed. The major QA items included the checks of the machine hardware, software and patient safety features. Briefly, these checks/tests include: 1) the cooling system reservoir and treatment table; 2) power to the treatment table; 3) the MR coil; 4) the transducer position with MRI; 5) image display on the treatment work station; 6) the effective focal spot in 3 directions using MR thermometry; and 7) all the safety devices including a sonication lamp, and the emergency stop‐sonication switches. In order to avoid patient skin burn, it is important to remove gas bubbles in the interfaces between the treatment table and the gel pad, and the gel pad and patients skin during the patient setup. Our QA procedures have been verified and evaluated through patient treatments. Seven patients with scapula, humeral head, sacrum, ilium, pubic ramus and acetabular bone metastases were treated using MRgFUS.Results:Our study showed that all seven patients tolerated the MRgFUS treatment well. No skin toxicity or other complications were observed. The pain score (0–10) using the visual analog scale (VAS) was significantly reduced from 8.0 ± 1.1 before treatment to 4.7 ± 3.0, 3.0 ± 1.5, 3.2 ± 2.8 and 3.4 ± 1.5 at one day, one month, two months and three months after the MRgFUS treatment, respectively.Conclusion:We demonstrated that with the appropriate QA procedures, MRgFUS is a safe, effective and noninvasive treatment modality for palliation of bone metastases.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call