Abstract

In this work, the results of the measurements of characteristics of 40-cm and 2-m scintillation counters irradiated by a particle beam with a momentum of 7 GeV/c from the accelerator of the Institute for High Energy Physics are presented. The scintillators used in counters are BC-404 and BC-408 scintillators. The counters are viewed from both ends by R1828-01 photomultiplier tubes. The PMT signal spectra are well described by a convolution of the Landau and Gaussian distributions. Their width is determined mainly by fluctuations of ionization energy losses. The time distributions of the signals obey the Gaussian law. For the 40-cm counter, the time resolution is σ(T) = 88 ps; for the 2-m counter, it varies from 120–160 ps in its center to ~100 ps near the end.