Abstract

ABSTRACT The surgical guide must be sterilized by standard process before clinical use. When the different sterilizations are processed, whether the surgical guide maintains the same accuracy creates doubts. Therefore, this study will delve into the changes of Dental SG resin in the 3D printing SLA process and MED 610 resin in the PolyJet process after sterilization in ahospital. The results show that the amount of change in the external deformation part of the Dental SG resin is about 0% to 0.6%, and the overall variation of the MED 610 resin is about 0% to 1.5%. In the internal stress analysis part, the amount of Dental SG resin before and after sterilization is about 2.35%~11.14%, and the change of MED 610 resin before and after sterilization is about 6.5%~9.05%. In this way, the changes are collected to provide abasis for the design of the posterior 3D printed surgical guide.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.