Abstract

The article is to study the objective evaluation method of the imaging quality of aspheric intraocular lens with different design concepts by establishing the model eye which is close to the actual corneal spherical aberration characteristics. Combined with the research data of corneal spherical aberration, the simulated cornea corresponding to Zernike c(4,0) coefficient 0.00 μm, 0.10 μm, 0.20 μm and 0.28 μm was established respectively. The optical quality of the measured simulated corneal lenses basically reached the theoretical design effect, and the deviation between the MTF value and the theoretical design value was equivalent to the MTF test repeatability. The establishment of model eye for artificial lens provides a research tool for the evaluation of aspheric intraocular lens imaging quality.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.