Abstract
Using the PGNAA set-up, which has been developed in our earlier work, a series of experiments are carried out to detect the boron and cadmium in aqueous solution. The minimum detectable concentrations of boron and cadmium are 7.0 and 2.5mg/L, respectively. But at the same time, it is found that the characteristic peak counts have poor linear relationship with the content of boron and cadmium. This will not only bring more error in the quantitative analysis, but also bring more difficulties to spectral resolution in the multi-element detection. In order to solve this problem, a correction factor g is proposed, and the experiment results have been corrected by using the g factor. The results shows that the linear relationship between the characteristic peak counts and the content of element have significantly improved after the correction. The linear correlation coefficients reach 0.99 or more.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.