Presented is a report on black-box evaluation of feedback control systems for commercial hearing aids. The aim of the study is to examine the ability of existing instrumental measures to quantify the performance of the feedback control system in black-box settings and on realistic signals, when more than one element of the signal processing chain may be active (compression, noise suppression, microphone directionality, etc.). The evaluation is carried out on 6 different hearing aids and for 10 measures. Thereby it is possible to see which measure is best suited to measuring which specific characteristic of the feedback control system, and serves as a beginning for conducting perceptual tests. The study uses static (but variable) feedback paths and is based on signals recorded from the in-ear microphone of an artificial head, on which the hearing instruments are mounted.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.