Abstract
Objective: To explore the effect of different post-processing parameters of digital radiography (DR) on the quality of chest X-ray for pneumoconiosis diagnosis, and to provide suggestions on parameter setting suitable for this kind of DR machine. Methods: From January 1, 2022 to June 30, 2022, the chest films of 35 workers in the department of radiology of Hangzhou occupational disease prevention and treatment hospital were randomly selected and printed after setting different image post-processing parameters. The quality of chest film was evaluated by the measurement of optical densitometer and the combination of subjective and objective by professional physicians. Results: When the density is set to 2 and the contrast/detail contrast is 4.5, the optical density of each area of DR chest film meets the requirements of chest X-ray quality, and the qualified rate of physician quality evaluation is the highest. Conclusion: Reasonable setting of image post-processing parameters can improve the quality of chest radiograph.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Zhonghua lao dong wei sheng zhi ye bing za zhi = Zhonghua laodong weisheng zhiyebing zazhi = Chinese journal of industrial hygiene and occupational diseases
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.