Abstract
The scintillator elements in X-ray CT equipment of the third generation type require high uniformity of distribution of light output as well as quality response characteristic. Parameters responsible for the distribution of light output were studied using the Monte Carlo method. Although the scattering coefficient was neglected, close agreement was obtained between calculated and experimental results. Calculated results indicated that relative output drop at both ends decreased with increasing optical absorption coefficient and was constant for X-ray absorption factor.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.