Abstract

The aim of this study was to evaluate noise exposure to the operating room staff consisting of the surgeon, assistant, anaesthetist and Mako Product Specialist (MPS) during Mako robotic-arm assisted total knee arthroplasty (TKA) and total hip arthroplasty (THA). We aimed to determine whether employees were exposed to noise at or above a lower exposure action value (LEAV) set out by the Noise at Work Regulations 2005, Health and Safety Executive (HSE), UK. We prospectively recorded intra-operative noise levels in Mako robotic-arm assisted TKA and THA over a period of two months using the MicW i436 connected to an iOS device (Apple), using the Sound Level Meter App (iOS) by the National Institute for Occupation Safety and Health (NIOSH). Data obtained was then used to calculate "worst case" daily exposure value to assess if sound levels were compliant with UK guidelines. Comparison between operating room staff groups was performed with ANOVA testing. A total of 19 TKA and 11 THA operations were recorded. During TKA, for the primary surgeon and the assistant, the equivalent continuous sound pressure level (LAeq) was over 80 dB, exceeding the LEAV set out by the Noise at Work Regulations by HSE. During THA, the average LAeq and peak sound pressure levels did not exceed the LEAV. The calculated daily exposure for the primary surgeon in TKA was 82 dB. A Tukey post hoc test revealed that LAeq was statistically significantly lower in the anaesthetist and MPS (p < .001) compared to the primary surgeon and assistant in both TKA and THA. Operating room staff, particularly the primary surgeon and assistant are exposed to significant levels of noise during Mako robotic-arm assisted TKA and THA. Formal assessments should be performed to further assess the risk of noise induced hearing loss in robotic-arm assisted arthroplasty.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call