Abstract
In the analysis of dioxins (PCDDs, PCDFs and DL-PCBs) whose isomers in total are more than 400 using gas chromatography-mass spectrometry, it is essential to design a suitable gas chromatographic condition to separate the peaks to be measured from other peaks. For this purpose, the literature was surveyed at first to know the separation ability of recent liquid phases for gas chromatographic columns, followed by characterization of experimental data obtained from three interlaboratory testings. A statistical calculation was applied to the results mentioned above to distinguish the separated peaks from the overlapped ones. Confident results were obtained after rejecting data of the overlapped peaks.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.