Abstract
The performance of our newly constructed cold-finger cryostat is investigated for the precision measurement of quantum Hall resistance (QHR) and single-electron pump (SEP) current. The cryostat is featured by a vacuum sealed sample chamber that is attached to the 0.3 K-plate inside the inner vacuum can. The nominal base temperature of the sample chamber was 0.4 K. However, the electron temperature estimated with a single-electron transistor was approximately 0.6 K. Precision evaluation of a GaAs QHR device shows that the quantized Hall resistance is equivalent to the nominal value of $h/2e^{2}$ within one part in 108. The Allan deviation data for the SEP current is analyzed to be greater than those obtained from SEP in a wet 3He cryostat at Korean Research Institute of Standards and Science.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.