Abstract

Crosslinked ultra-high molecular weight polyethylene (UHMWPE) is a preferred material used as a bearing material in the artificial hip joint. The purpose of this study was to analyze the effect of dose cross-linking acetabular liner manufactured by CNC milling on wear depth. The wear test is carried out with a tribometer machine by loading the acetabular liner in a certain cycle. The test results found that specimens given a crosslink dose of 50 kGy with a load of 800 N resulted in a greater wear depth compared to specimens given a crosslink dose of 100 kGy with a load of 800 N. Specimen with crosslink dose 50 kGy yield greater wear depth than specimens given a crosslink dose of 100 kGy. Based on these results it can be concluded that the greater the crosslink dose given to the specimen will increase the wear resistance of the specimen so that the test results in a smaller wear depth.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.