Abstract

In ultrasonic flow metering, the inner pipe wall will be distorted by transducer’s installation, and the velocity profile of the meter will be different from that in a long straight pipe located in front of or behind the meter section. Changes in profile will finally influence the flow measurement accuracy. This paper shows a multi-path ultrasonic flow meter with its transducers installed at different positions. Our purpose is to find the best transducer installation position and to study the influence of path numbers and pipe diameters on the flow measurement accuracy at the best installation position. With a computational fluid dynamics (CFD) analysis, velocity profiles are calculated. Negative bias errors are obtained at all the three of the common installation positions. Similar phenomenon has also been observed in a real field experiment. The reasons for the negative bias error have also been analyzed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.