Abstract

Optimation of simulation design of collimator is corresponding to 30 MeV cyclotron generator. The simulation has used the variation of the thickness materials layers that was applied at treatment room’s door. The purpose of the variation and thickness of the material in this simulation to obtain optimum results for the shielding design in the irradiation chamber. The layers that we used are Pb-Fe and Pb-SS312. Simulation on cancer treatment is used with monte carlo simaulation MCNPX. The spesifications that we used for cyclotron is the spesification of the HM-30 Proton Cyclotron from Sumitomo Heavy Industries Ltd. The variation of the thickness materials layers that was applied at treatment room’s door are Pb remains 4cm while Fe and SS312 varies between 2 cm, 4 cm, 6 cm respectively. This simulation of Fe layer on Pb was give good result in measurement simulation at 4 cm thickness.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.