Abstract

The change in light intensity that takes place when an ambient air sample is drawn into the detection chamber of a chemiluminescence monitor generates changes in the concentrations of several species, such as NO(2), NO and O(3). Although this phenomenon has been known for several decades, there is still no commonly accepted approach on when or how to correct for it in NO(2) and O(3) readings. In this work, we have assessed the expanded uncertainty of two chemiluminescence NO( x ) analysers commercially available according to EN 14211:2005, with the aim of establishing the maximum allowable standard uncertainty due to the reaction between NO and O(3) in the sampling system. Although this maximum allowable uncertainty cannot be a universal value-as it will depend on the performance of each analyser-our results have led us to propose the conservative value of 2%. We have also proposed a methodology for improving data quality which could be easily implemented by those responsible for air quality data validation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.