Abstract

A common problem shared by accelerometers, inertial sensors and any motion measurement method based on skin-mounted sensors is the movement of the soft tissues covering the bones. The aim of this work is to propose a method for the validation of the attachment of skin-mounted sensors. A second-order (mass–spring–damper) model was proposed to characterize the behaviour of the soft tissue between the bone and the sensor. Three sets of experiments were performed. In the first one, different procedures to excite the system were evaluated to select an adequate excitation stimulus. In the second one, the selected stimulus was applied under varying attachment conditions while the third experiment was used to test the model. The heel drop was chosen as the excitation method because it showed lower variability and could discriminate between different attachment conditions. There was, in agreement with the model, a trend to increase the natural frequency of the system with decreasing accelerometer mass. An important result is the development of a standard procedure to test the bandwidth of skin-mounted inertial sensors, such as accelerometers mounted on the skin or markers heavier than a few grams.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.