Abstract

The europium–thenoyltrifluoroacetone (TTA)–cetyltrimethylammonium bromide (CTAB)–Triton X-100 system at pH 5.5–7.5 was studied and used for the spectrofluorimetric determination of europium. The excitation and emission wavelengths were 370 and 612 nm, respectively. Fluorescence intensity was a linear function of the concentration of europium in the range 1.0 × 10–7–1.0 × 10–9 mol dm–3. The detection limit was 2.0 × 10–12 mol dm–3. The standard additions method was used for the determination of europium in synthetic rare earth oxides and gave satisfactory results. The molar ratio method was carried out spectrofluorimetrically for the determination of the stoichiometric composition of the fluorescent species. A ratio of 1 : 4 : 1 for the ternary Eu–TTA–CTAB complex was found.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.