Abstract
The study on the effective radii of ultrasonic transducers for hydrophone calibration at the National State Primary Standard of the Unit of Ultrasonic Pressure in Water (NDETU AUV-02-2018) is presented and the appropriate method is described. The impact of the effective radius on measurement distance and diffraction loss coefficients is evaluated. The uncertainty calculation of the effective radius measurement of ultrasonic transducers is provided, and its influence on the accuracy of hydrophone calibration is assessed. Considerable attention is given to estimating the measurement distance between the hydrophone and the ultrasonic transducer, which significantly affects the accuracy of hydrophone calibration.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.