Abstract
A study was undertaken to determine whether involuntary user movement provides a basis for relaxing the measurement conditions for evaluating the potential optical radiation hazards to the eye from slit lamps and indirect ophthalmoscopes. This was accomplished by assessment of the extent to which light from these devices can be maintained in focus on a 1-mm-diameter fiber-optic cable for 45 s. The results suggest that, although involuntary user movements can be significant, they do not provide a basis for relaxing the measurement conditions for evaluating the potential optical radiation hazards to the cornea and lens from slit lamps and indirect ophthalmoscopes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.