Abstract

In the present contribution, sensitive and precise method for the quantification of Organophosphorus / Pesticides (Malathion and Dimethoate) in nanograms range has been developed. The performance of flame photometric detector (FPD), a selective detector (P&S-mode) that can be used in the analysis of organophosphorus compound, is evaluated in terms of sensitivity, selectivity and reproducibility. The performance of flame photometric detector was strongly depending on the absolute and relative flow rate of air and hydrogen gases. The optimum air-to-fuel ratio for detection of Malathion and Dimethoate was 0.4 and 0.3 (FPD-P mode). At this ratio, low picogram amounts of phosphor can be detected accurately (0.18 pgP) with a wide linear dynamic range of 0.18 pgP to 298 ngP. While, the optimum air-to-fuel ratio, for detection of Malathion and Dimethoate was 0.6 (FPD-S mode). In addition to, the method is precise with 4.5 % relative standard deviation (RSD). In conclusion, it could be proposed that this procedure can be recommended as a suitable method for the quantification of Malathion and Dimethoate in cases of acute poisoning.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.