Abstract

One of the major obstacles facing radioimmunotherapy (RIT) is the lack of a device to measure directly tumor and normal tissue radiation absorbed dose. Calculations based on the clearance and imaging scans have several limitations; hence we design and fabricate a sheathed clinical thermoluminescent dosimeter (TLD) for the measurement of absorbed dose by implantation in humans. Preclinical studies are performed in nine normal rabbits. Complete blood count, body temperature monitoring, clinical observation and necropsy show no untoward effects from the TLD. Consistent bone marrow radiation doses are noted in the four rabbits receiving 131I-labeled monoclonal antibody A6H. By using up to 20 clinical TLDs in one sheath, it will be possible to determine macroscopic heterogeneities in organs undergoing RIT.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.