Abstract

This is a first report on a new promising red-region fluorescence substrate, tetra-substituted amino aluminum phthalocyanine, which displays an excitation maximum at 610 nm and an emission maximum at 678 nm in strongly acidic medium. It has been synthesized and applied to the determination of traces of hydrogen peroxide and horseradish peroxidase (HRP). The steady-state catalytic rate depends upon the enzyme and substrate concentrations, and the Michaelis–Menten parameters K m and V max are measured to be 2.82×10 −6 mol l −1 and 6.0×10 −9 mol l −1 s −1, respectively. Under optimum conditions, the calibration graph has a linear range of 0.0 to 3.94×10 −11 mol l −1 HRP and 0.0 to 2.0×10 −7 mol l −1 hydrogen peroxide, with 3 σ detection limits of 5.9×10 −13 mol l −1 HRP and 1.4×10 −9 mol l −1 H 2O 2. By coupling with a glucose oxidase-catalyzed reaction, glucose in human serum has been quantified and the results are in good agreement with those reported by a hospital laboratory. The proposed method can greatly decrease the interference that results from background fluorescence or scattered light and has a high analytical sensitivity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.