Abstract
Linear variable filters (LVF) are widely used in a variety of small rapid spectrometric detecting devices. In the present paper, the splitting property of LVF was studied, and the spectral-property expression in Gaussian form was given, and the relationship between indexes of expression and LVF center transmission, width of spectrum, and linear dispersion was specified. Measurement based on monochromator for imaging spectrometers calibration was introduced, the sensitivity for detection system was discussed, and the tolerance was analyzed. It was shown that the slit displacement of monochromator relative to the optical axle and the slope of LVF affect the detecting accuracy most, and the tolerance can be reduced by adjustment of optical path and structure to reach requirement. System was built and the detection for spectral property of sample LVF was made. The result of data processing indicates that MSE for center transmission detection was lower than 0.05%, and the accuracy of this method was proved. Reference-parameters of relevant system designing and calibrating based on the LVF can be provided by this method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.