Abstract

Mammography techniques have recently advanced from those using analog systems (the screen-film system) to those using digital systems; for example, computed radiography (CR) and flat-panel detectors (FPDs) are nowadays used in mammography. Further, phase contrast mammography (PCM)-a digital technique by which images with a magnification of 1.75× can be obtained-is now available in the market. We studied the effect of the air gap in PCM and evaluated the effectiveness of an antiscatter x-ray grid in conventional mammography (CM) by measuring the scatter fraction ratio (SFR) and relative signal-to-noise ratio (rSNR) and comparing them between PCM and the digital CM. The results indicated that the SFRs for the CM images obtained with a grid were the lowest and that these ratios were almost the same as those for the PCM images. In contrast, the rSNRs for the PCM images were the highest, which means that the scattering of x-rays was sufficiently reduced by the air gap without the loss of primary x-rays.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.