Abstract
Many modern molecular imaging techniques, based on radiopharmaceuticals, can take advantage of sophisticated devices but are still based on the scintillation detector mechanism proposed by Anger. These devices can perform with millimeter spatial resolution and high detection efficiency, but the final performance is strongly affected by the algorithm used for the scintillation position detection.In this work, a detailed comparison of the effect on the imaging performances of three new position detection algorithms, in terms of spatial resolution, detection linearity and useful Field of View is performed on a prototype gamma detector. The detector, built by the authors, is based on a continuous LaBr3:Ce scintillation crystal coupled to an Hamamatsu MA-PMT H8500 and a single anode readout electronics. The experimental data are obtained scanning the detector surface with a Tc99m collimated source (0.4 mm ϕ) at 1.5 mm step.The overall imaging performances of the device are also tested by mean of a bar phantom. We conclude that the Anger-like algorithms give a 50% uFoV with a 1.30 mm ±0.05 mm spatial resolution while the proposed algorithms give a 80% uFoV and 1.10 mm ±0.06 mm spatial resolution.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.