Abstract

Two magnetic field monitoring systems for g-2/EDM experiment at J-PARC are developed. One is a continuous-wave NMR magnetometer, which is used to measure the field homogeneity in a muon storage region. The other is a Hall effect magnetometer, which is used to measure the magnetic field around the muon injection region and that for focusing the muon beam. In order to improve the accuracy of the commercial NMR magnetometer, a digital signal processing system was implemented instead of an analog one. As for the Hall probe system, a three-axis Helmholtz coil system was fabricated to measure tilt angles of Hall sensors in a three-axis Hall probe. The field homogeneity of each Helmholtz coil was measured, and the result agreed well with the design values.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.