Abstract

Levofloxacin (LEV) is a type of fluoroquinolone antibiotic that usually used for treating the bacterial infection. The released of LEV in environment may impact a significant risk to the ecosystems. Thus, a fast and sensitive sensor device is required. In this work, the detection of LEV is carried out using a screen-printed carbon electrode (SPE). The measurement methods used were square wave voltammetry and cyclic voltammetry. The limit of detection and limit quantitation were 4.34 µM, 14.4 µM, respectively. The relative standard deviation was obtained at 5.4%. The %recovery results obtained using screen printed electrode in drug, milk, and wastewater were in the range of 95-110%. The validated method was successfully applied to detect the levofloxacin and resulted in a sensitive and efficient measurement.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.