Abstract
This research presents an experimental study evaluating stomach suturing using a precurved nickel-titanium (NiTi) guidewire for an endoscopic minimally invasive obesity treatment. Precise path planning is critical for accurate and effective suturing. A position measurement system utilizing a hand-held magnetic sensor was used to measure the shape of a precurved guidewire and to determine the radius of curvature before and after suturing. Ex vivo stomach suturing experiments using four different guidewire tip designs varying the radius of curvature and bevel angles were conducted. The changes in radius of curvature and suturing force during suturing were measured. A model was developed to predict the guidewire radius of curvature based on the measured suturing force. Results show that a small bevel angle and a large radius of curvature reduce the suturing force and the combination of small bevel angle and small radius of curvature can maintain the shape of guidewire for accurate suturing.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.