Abstract

Hemolysis can significantly affect the reliability of test results and occur in the pre-analytical phase. The aim of this study is to reveals the correlation of hemoglobin levels on hemolysis sample. This experimental study was carried out using samples from thirty students of Medical Laboratory Technology study program of the Nahdlatul Ulama University of Surabaya. Blood samples were treated by hard shaken in 30 times in order to damage the middle part of the blood sample. Data on hemoglobin levels were collected and analyzed using the Pearson Correlation Test before and after treatment with significant value p < 0.05 indicating that there was a significant correlation. Hemoglobin concentrations were strongly positive for the hemolysis of the sample (p = 0.000). The conclusion is that sample hemolysis has a potency to be the confounding factor on the hemoglobin test.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.