Abstract
It is a useful method for the adaptive radiotherapy (ART) to calculate absorbed dose accurately on the image set taken by on-board cone beam computed tomography (CBCT) attached to linac for image-guided radiation therapy (IGRT). For this purpose, a simple and accurate calculation method is necessary. Several papers report that it is possible to calculate easily and accurately by using several methods of researches in the neck and prostate, but the lung density varies greatly depending on patient thorax condition. In this study, we propose a new dose recalculation method, which is a simple procedure and can achieve accurate dose calculation considered different lung densities in each patient. By using this method, it is possible to calculate exclusive of artifacts in CBCT because of overriding the lung density. The dose error between dose recalculation of the CBCT image and treatment plan agreed within±1%. Therefore, this method is expected to be a useful method for accurate dose calculation with CBCT image for ART.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.