Abstract
A low energy secondary ion mass spectrometer was constructed. An argon ion gun was designed by the authors and a modefied quadrupole type mass analyser was used as a detector. The best operating condition was established as follows: ion accelerating voltage; 500V, argon pressure;5×16mm. Surface contamination on copper surface with abrasive materials were investigated for chemical etching with hydrochloric acid Al2O3powder emulsion polishing, Al2O3abrasing paper polishing and SiC abrasing paper polishing. The latter two methods gave very strong peak of27Al. The best result for positive ions was obtained with samples treated by chemical etching with hydrochloric acid.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.