Abstract

The aim of this study was to investigate the effect of ageing and fading on PADC detector response, as a function of the storage time and temperature. Several groups of CR-39 detectors provided by Radosys, Ltd. were exposed at the reference radon chamber of the Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti, at the ENEA centre in Casaccia. The results indicate that low-temperature storage inhibits the effect of both ageing and fading. Finally, the overall reduction in CR-39 sensitivity due to the combined ageing/fading effect was estimated. In particular, the sensitivity of the detectors continuously exposed in air at room temperature over 6 and 3 months was reduced, respectively, by 7.5 and 4 %.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.