Abstract

Aqueous sample solutions were nebulized by an ultrasonic nebulizer and the solvent was removed by a desolvation apparatus before introduction of the aerosol into the microwave-induced plasma (MIP). The desolvation system consists of a heating tube and a condenser associated with a concentrated H2SO4 desiccator. The water vapor was removed almost completely with the desolvation system. The detectability of microwave-induced plasma atomic emission spectrometry was therefore improved. The detection limits for P, Cl, Br, and I were 0.0045, 0.12, 0.23. and 0.06 μg/ml, respectively. Based on a study of the emission spectra of P, Cl, Br, and I in the range of 200 to 600 nm in He MIP, the optimum analytical lines for determination of corresponding elements were chosen.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.