Abstract

Some results of a study of the properties of a low pressure multistep avalanche chamber with BaF 2 scintillators and with TMAE as the photosensitive gas for a 3D PET scanner are presented. Absolute gas amplification was carefully measured by two independent methods, giving rise to a maximum attainable total gain of the order of 10 7. A typical electron transmission efficiency of 20–30% was obtained. The time resolution of the detector can be optimized to be ∼ 6 ns FWHM.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.