Abstract
An accurate method for measuring charge transfer efficiency (η) and photoresponse nonuniformity (PRNU) of linear charge-coupled devices is developed. The apparatus is described. A finely focused light spot is adopted to irradiate each photosensitive element to get its response output signal. Data are collected automatically by scanning peak-searching and processed by linear best fitting. As examples, measurements for some types of devices are reported and factors influencing the results are discussed.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.