Abstract

An accurate method for measuring charge transfer efficiency (η) and photoresponse nonuniformity (PRNU) of linear charge-coupled devices is developed. The apparatus is described. A finely focused light spot is adopted to irradiate each photosensitive element to get its response output signal. Data are collected automatically by scanning peak-searching and processed by linear best fitting. As examples, measurements for some types of devices are reported and factors influencing the results are discussed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.