Abstract

(1) In order to find an adequate lead for recording the electrocardiogram during exercise, 6 modifications of Frank's Y-lead were studied.(2) For this purpose, factors examined were the phase difference, the voltage, the wave form as cardiac factors and the interference by electrical activity of the muscle and the influence of respiration as non cardiac factors.(3) The study in 34 cases without the marked intraventricular conduction disturbance indicated that the lead, in which a positive electrode at the height of the umbilicus and a negative one is placed in the subclavicular region along the right mid-clavicular line, is recommended.(4) In practice, electrodes by Frank's lead system and other 2 extraelectrodes by the modified lead are placed. The amplification of the modified lead is adjusted to the amplitude of the original one. With this method, records which is electrically equivalent to those by Frank's lead system can be taken with little interference.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.