Abstract

One of the most important characteristics of thermoluminescent detectors, if they are intended to be used in a wide range of energies, is their energy response. A comparison of the energy characteristics of the newly developed LiF:Mg,Cu,P chips with additional PbO doping (GR-200E) against GR-200A LiF:Mg,Cu,P was attempted to improve further the accuracy of personal dosimetry. Hp(10) energy response of GR-200E is quite different from that of GR-200A. For GR-200E, the anomalous energy response can be eliminated and the over-response to low-energy photons should be filtered properly. The GR-200E under the copper filter provided results within 4 % with respect to the conventional quantity value over the energy range from 65 to 1250 keV. The ratio of the Hp(10) response of the Cu-covered filters for GR-200A to that of those covered with plastic for GR-200E is a very good tool for identifying the radiation quality in the range studied. The accuracy in the dose evaluated from TLD measurements can be greatly increased.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.