Abstract

A new method has been developed for ion-interaction chromatography with suppressed conductivity detection and a new graphitized carbon packing, which is sintered from carbonic material at a high temperature. Combinations of various eluting agents, tetrabutylammonium hydroxide (TBA) and acetonitrile have been investigated to optimize the separation of eight common anions (F −, Cl −, NO 2 −, Br −, NO 3 −, SO 4 2−, HPO 4 2− and I −). Calibration curves were linear from 0.5 to 10 μg/ml for F −, from 1.0 to 20 μg/ml for Cl −, NO 2 − and NO 3 −, from 2.5 to 50 μg/ml for Br − and SO 4 2− and from 5.0 to 100 μg/ml for HPO 4 2− and I − with a correlation coefficient ( r) of 0.999 or better. The relative standard deviations (R.S.D.s) of peak areas were between 0.2 and 0.9% for 10 repeated measurements. The application of this newly developed method was demonstrated by the determination of chloride, bromide and sulfate in pharmaceutical compounds using the direct injection method. The analytical results were within ±2% (relative) of the theoretical value, and thus in good agreement with the theoretical value for each sample.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.