Abstract

The results from studies of the effects of several factors on the accurate measurement of particulate matter (PM) mass collected on ambient air monitoring filters are described. The main investigation quantifies the effect of humidity on the mass of blank filters of some commonly used materials in Europe, namely quartz fibre, glass fibre and PTFE-bonded glass fibre. Supplementary work investigates the effect of humidity on the mass of sampled PM, and effects of storage and transport are also reported. A number of fundamental problems associated with the weighing of filters are quantified and, for the materials and factors studied, we conclude that quartz filters are less well suited to high accuracy determinations of PM mass than glass fibre filters, while PTFE-bonded glass fibre filters are better than both. Though other factors also have an important role in the choice of filter material, we believe that the results form a useful addition to the limited amount of information available in this area.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.