Abstract

Abstract Experiments with a simple bridge for the determination of polarization resistance, electrode capacitance, and time constant of corroding electrodes are presented. The novelty is the use of low-frequency square waves (e.g., 10 mHz, 5 mV peak-to-peak) impressed on the specimens via capacitors in parallel with resistors. The instrumentation is very cheap and can be put together easily. The method is non-disturbing and reasonably fast. Data from eight different corroding systems have been compared with results from electrochemical impedance spectroscopy (EIS) and potentiokinetic sweep. The polarization current can alter performance of electrodes, so polarization resistance varied substantially with the type of measurements and the measuring parameters for seven out of the eight test cases. Therefore, the instrument was designed to handle modeling with two resistance-capacitance (RC) links in series. Since the lack of a well-defined polarization resistance is not easily observed with conventional meth...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.