Abstract

Several classes of environmentally significant compounds were introduced into an ion trap detector (ITD) by capillary column gas chromatography, and the performance of the device as a gas chromatography/mass spectrometry (GC/MS) system was studied. The experiments conducted included studies of the precision of selected ion monitoring (SIM) measurements, the precision of relative ion abundance measurements in full mass range scans, the sensitivity and instrument detection limits in SIM and full mass range data acquisition, and the linear range of concentration calibration. It was observed that resolution may be lost and saturation may occur under certain conditions, and that the data system can report artifact ions under these conditions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.