Abstract

Structured illumination microscopy applied to in-vivo retinal imaging has the potential to provide a low-cost and powerful diagnostic tool for retinal disease. In this paper the key parameters that affect performance in structured illumination ophthalmoscopy are studied theoretically. These include the number of images that need to be acquired in order to generate a sectioned image, which is affected by the non-stationary nature of the retina during acquisition, the choice of spatial frequency of the illuminating sinusoid, the effect of typical ocular aberrations on axial resolution and the nature of the sinusoidal pattern produced by the illumination system. The results indicate that structured illumination ophthalmoscopy can be a robust technique for achieving axial sectioning in retinal imaging without the need for complex optical systems.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.